Teixeira, C. A.Ruano, AntonioRuano, M. GraçaPereira, W. C. A.Negreira, C.2013-02-072013-02-072006Teixeira, C. A.; Ruano, A. E.; Ruano, M. Graça; Pereira, W. C. A.; Negreira, C. Non-invasive temperature prediction of in vitro therapeutic ultrasound signals using neural networks, Medical & Biological Engineering & Computing, 44, 1-2, 111-116, 2006.0140-0118AUT: ARU00698; MRU00118;http://hdl.handle.net/10400.1/2251In this paper, a novel black-box modelling scheme applied to non-invasive temperature prediction in a homogeneous medium subjected to therapeutic ultrasound is presented. It is assumed that the temperature in a point of the medium is non-linearly related to some spectral features and one temporal feature, extracted from the collected RF-lines. The black-box models used are radial basis functions neural networks (RBFNNs), where the best-fitted models were selected from the space of model structures using a genetic multiobjective strategy. The best-fitted predictive model presents a maximum absolute error less than 0.4 C in a prediction horizon of approximately 2 h, in an unseen data sequence. This work demonstrates that this type of black-box model is well-suited for punctual and noninvasive temperature estimation, achieving, for a single point estimation, better results than the ones presented in the literature, encouraging research on multi-point non-invasive temperature estimation.engUltrasoundHyperthermiaNeural networkNon-invasive temperature prediction of in vitro therapeutic ultrasound signals using neural networksjournal article2013-01-26http://dx.doi.org/10.1007/s11517-005-0004-2