This work reports the assessment of time-shifts (TS) from backscattered ultrasound (BSU) signals when large temperature variations (up to 15 degrees C) were induced in a gel-based phantom. The results showed that during cooling temperature is linear with TS at a rate of approximately 74 ns/degrees C. However during a complete heating/cooling cycle, the relation is highly non-linear. This can be explained by the fact that during cooling the temperature distribution is more uniform. Another problem to report is that TS is very sensitive to external movements.
Description
Keywords
Non-invasive temperature estimation Temperature models Time-shift estimation Ultrasound